タイトル |
-
en
Application of an ultraminiature thermal neutron monitor for irradiation field study of accelerator-based neutron capture therapy
|
作成者 |
|
アクセス権 |
open access |
権利情報 |
|
主題 |
-
Other
en
neutron capture therapy
-
Other
en
SOF detector
-
Other
en
thermal neutron monitor
|
内容注記 |
-
Abstract
en
Phantom experiments to evaluate thermal neutron flux distribution were performed using the Scintillator with Optical Fiber (SOF) detector, which was developed as a thermal neutron monitor during boron neutron capture therapy (BNCT) irradiation. Compared with the gold wire activation method and Monte Carlo N-particle (MCNP) calculations, it was confirmed that the SOF detector is capable of measuring thermal neutron flux as low as 105 n/cm2/s with sufficient accuracy. The SOF detector will be useful for phantom experiments with BNCT neutron fields from low-current accelerator-based neutron sources.
|
出版者 |
en
Oxford University Press
|
日付 |
|
言語 |
|
資源タイプ |
journal article |
出版タイプ |
VoR |
資源識別子 |
HDL
http://hdl.handle.net/2115/58566
|
関連 |
-
isIdenticalTo
DOI
https://doi.org/10.1093/jrr/rru112
-
PMID
25589504
|
収録誌情報 |
-
-
PISSN
0449-3060
-
EISSN
1349-9157
-
en
Journal of Radiation Research
-
巻56
号2
開始ページ391
終了ページ396
|
ファイル |
|
コンテンツ更新日時 |
2023-07-26 |